Featuring These Great Additions Over Previous Models:
Automatic, menu-driven operation requires minimal user skills
Greater detection range, from 0.005% - 100.0%
Greater temperature range, programmable start and test temperatures from 25- 275 °C
Storage of up to 250 test programs, for single-button, product-customized testing
User-programmable calibration prompts for simple balance and temperature calibration
Self-diagnostics alert operator of any instrument problems
Lift compensation - compensates for pan lift caused by convection currents in moisture analyzers utilizing Loss-on-Drying technology>
Security feature restricts access to program and instrument settings
Optional multi-point temperature calibration module (utilizes NIST-traceabled RTD calibrated to ±1 °C)
Optional validation package in Adobe Acrobat format and validation services available
MAX 2000XL Specifications:
Sample Size 150 mg to 40 g
Resolution Moisture: 0.001%; Balance: 0.0001g
Moisture/Solids Range 0.005% to 100%
Dry Weight Range 0-300%
Results Display % Moisture, % Solids, or % Dry Weight
Ending Criteria Predict (Auto or Manual), Rate, or Time
Heating Range 25 - 275 °C, Set in 1 °C increments and maintained to ±1 °C within the range of 100 - 200 °C
Test Parameter Memory 250 Programs
Program Identification Alphanumeric and/or symbols; programmable prompts for lot or product ID numbers
Test Result Memory Data from last 108 tests
Statistical Calculation Mean, Standard Deviation, and Coefficient of Variation
Power Requirements 100-120V ~ 50/60 Hz, 1 amp (standby) or 8 amps (heat on),
or 220-240V ~ 50/60 Hz, 0.5 amp (standby) or 4 amps (heat on)
Interfaces RS-232 (9-pin standard); parallel printer (LPT - PC standard)
Dimensions 235mm H x 300mm W x 575mm D; 9.5" H x 12" W x 24" D
Weight 13.9 kg; 32 pounds
Warranty One years
Certifications U.S. and Canadian Underwriters Laboratory; European Communities - CE
Accessories Multi-Point Temperature Calibration Module; Printer; Nitrogen Purge Fittings; Carrying Case; Sample Basket; Validation Package; Bi-directional Communication Package; Data Collection Package
Additional Features:
User-programmable target sample weight from 150mg to 40g eliminates test variables
Easy setup of product methods to ensure optimum test results
Starting temperatures are programmable and maintained within ± 1 °C increments
Camped temperature programs for quicker test times, and linked temperature programs for determining free and bound moistures
Real-time graph of test’s drying curve
Flexible ending criteria to ensure accurate test results
Stores over 100 tests and statistical features allow calculation of mean, mean deviation and coefficient of variation
Nitrogen purge minimizes oxidation of samples and flashing of volatiles
Syringe function eliminates solvent evaporation and captures accurate start weight
Foil weight entry mode for paint solids at OEM assembly plants
Optional bi-directional communication package for data storage and retrieval from PC and easy programming of multiple instruments
Optional data collection package for downloading results to your PC