Time Base Characteristics |
Type | TCO (Temperature Controlled Oscillator) |
Line Voltage Stability | Less than ± 1 PPM with ± 10% line voltage variation |
Int. Std. Out | 10 MHz (Internal Standard Frequency Output) |
Level | 1 V p-p or more. TTL (10TTL) |
Ext. Std. In | 10 MHz (External Standard Frequency Input) |
Display | Nine Digit 0.56" LED with M (Mega) or K (Kilo) Hz; nano, micro, or milli (n/µ/m) Sec.; GT (GATE TIME); H (HOLD); and OF (OVERFLOW) indicators. FUNCTION and GATE TIME are user selected. OF display is shown when the count exceeds 999 999 999 |
Hold | In the FREQUENCY, PERIOD, TOTAL, RPM modes, measurement in progress is stopped; and the last complete measurement is displayed. When HOLD is released, a new measurement begins |
Gate Time | Depending on input frequency < 10 mS - Somewhere between 0.9 and 9mS < 0.1 S - Somewhere between 9 and 90mS < 1 S - Somewhere between 90 and 900mS < 10 S - Somewhere between 0.9 and 9S |
Note | Last measurement display will remain for 10 seconds after signal off |
Dimensions (WxHxD) | 9.4 x 3.5 x 10.6" (240 x 90 x 270mm) |
Power | 115VAC±10%, 50-60 Hz, 9 W |
Plug and Socket | 3 wire ac power plug and 3 wire outlet |
Temperature | 0 °C to +40 °C (Accuracy Specified at 23 °C ± 5 °C) |
Humidity | up to 85% RH (Relative Humidity) to 40 °C without temperature extremes causing condensation within the instrument |
Insulation Category II | Portable equipment of local level |