The THDP0100 and P5210A have the largest differential dynamic range capability from Tektronix, allowing users to safely measure up to ±6000 V. The THDP0100 supports bandwidths up to 100 MHz and slew rates up to 2500 V/ns at 1/1000 gain. These probes
Characteristic | TMDP0200 | THDP0200 | THDP0100 |
Attenuation | 25X / 250X | 50X / 500X | 100X / 1000X |
Differential Voltage | 250X: ±750 V 25X: ±75 V | 500X: ±1500 V 50X: ±150 V | 1000X: ±6000 V 100X: ±600 V |
Common Mode Voltage | ±750 V | ±1500 V | ±6000 V |
Maximum Input Voltage-to-Earth | 550 V CAT I 300 V CAT III | 1000 V CAT II 600 V CAT III | 2300 V CAT I 1000 V CAT III |
Bandwidth | 200 MHz | 200 MHz | 100 MHz |
Rise Time | <1.8 ns | <1.8 ns | <3.5 ns |
Slew Rate | <275 V/ns at 1/250 gain | <650 V/ns at 1/500 gain | <2500 V/ns at 1/1000 gain |
Input Impedance at the Probe Tip | 5 MΩ || <2 pF | 10 MΩ || <2 pF | 40 MΩ || <2.5 pF |
Typical CMRR | DC: > –80 dB 1 MHz: > –60 dB 3.2 MHz: > –30 dB 50 MHz: > –26 dB | DC: > –80 dB 1 MHz: > –60 dB 3.2 MHz: > –30 dB 50 MHz: > –26 dB | DC: > –80 dB 1 MHz: > –60 dB 3.2 MHz: > –30 dB 100 MHz: > –26 dB |
Cable Length | 1.5 m | 1.5 m | 1.5 m |
Probes with the TekProbe® Interface
Characteristic | P5200A | P5202A | P520 | P5210A |
Attenuation | 50X / 500X | 20X / 200X | 50X / 500X | 100X / 1000X |
Differential Voltage | 500X: ±1300 V 50X: ±130 V | 200X: ±640 V 20X: ±64 V | 500X: ±1300 V 50X: ±130 V | 1000X: ±5600 V 100X: ±560 V |
Common Mode Voltage | ±1300 V | ±640 V | ±1300 V | ±5600 V |
Maximum Input Voltage-to-Earth | 1000 V CAT II | 450 V CAT l 300 V CAT ll | 1000 V CAT II | 2300 V CAT l 1000 V CAT III |
Bandwidth | 50 MHz | 100 MHz | 100 MHz | 50 MHz |
Differential Input Impedance | 10 MΩ, 2 pF | 5 MΩ, 2 pF | 10 MΩ, 2 pF | 40 MΩ, 2.5 pF |
Input Impedance between each Input and Ground | 5 MΩ, 4 pF | 2.5 MΩ, 4 pF | 5 MΩ, 4 pF | 20 MΩ, 5 pF |
Typical CMRR | DC: >80 dB 100 kHz: >60 dB 3.2 MHz: >30 dB 50 MHz: >26 dB | DC: >80 dB 100 kHz: >60 dB 3.2 MHz: >40 dB 50 MHz: >30 dB |
Cable Length | 1.8 m |
1. This is the maximum measurable range between the (+) and (–) input leads of the probe. Beyond these limits, the output could be clipped.2. The maximum common mode voltage and maximum input voltage-to-earth (RMS) are the maximum voltages that each input lead (+/–) can be from ground.