Bridge power supply | DC10 V±0.3 V within 120 mA (Changeable to DC2.5 V or 5 V) with remote sensing function |
Input range | F.S. setting can be available at the input from 0.3 mV/V to 3 mV/V. (when bridge power supply is DC10 V) |
Zero adjustment range | 0 mV/V to 2.4 mV/V |
Non-linearity | 0.01 %F.S. |
Temperature coefficient | Zero point ±0.2 μV/℃ (Input conversion) Sensitivity ±0.001 5 %F.S./℃ |
Load display | Display of gross weight and net weight |
Display range | -1 000 to 11 000 (Changeable to x2, x5, x10, x20 or x50) |
Display range | FINAL : 5 digits PRE 1 : 5 digits PRE 2 : 5 digits FALL : 3 digits OVER : 2 digits UNDER : 2 digits CODE : 2 digits(16 kinds of settings are available.) |
Condition display | |
arrow section light on | During indication of Zero point, Stable condition, Gross weight, Tare Cancellation, SP1, SP2 and SP3. |
Description display | Display of the set value for FINAL, PRE 1, PRE 2, FALL, OVER, UNDER, CODE and SETTING. |
Unit | kg, t (LED light on) |
Display times | 20 times/s (Changeable to 4 times/s) |
Operating temp./humidity range | Temperature -10 ℃ to 50 ℃, Humidity 85 % or less (Non condensing) |
Power supply voltage | AC100 V(AC85 V to AC100 V) 50/60 Hz |
Power consumption | Approx. 25 VA |
Outline dimensions(WxHxD) | 192 mm x 96 mm x 175 mm (excludes protruding parts) |
Weight | Approx. 2.5 kg (without options) |