Vertical System Analog Channels |
Input Channels | Analog Bandwidth (–3 dB) | 350 MHz |
Calculated Rise Time | 1 ns |
DC Gain Accuracy | ±1.5%, derated at 0.10%/°C above 30°C |
Bandwidth Limits | Depending on instrument model: 1GHz, 500MHz, 350MHz, 250MHz, and 20MHz |
Input Coupling | AC, DC |
Input Impedance | 1 MΩ ±1%, 50 Ω ±1% |
Input Sensitivity | 1 MΩ: 1 mV/div to 10 V/div 50 Ω: 1 mV/div to 1 V/div |
Vertical Resolution | 8 bits (11 bits with Hi Res) |
Max Input Voltage, 1 MΩ | 300 VRMS CAT II, with peaks ≤ ±425 V For <100 mV/div derate at 20 dB/decade above 100 kHz to 30 VRMS at 1 MHz, 10 dB/decade above 1 MHz For ≥100 mV/div derate at 20 dB/decade above 3 MHz to 30 VRMS at 30 MHz, 10 dB/decade above 30 MHz |
Max Input Voltage, 50 Ω | 5 VRMS, with peaks ≤ ±20 V |
Position Range | ±5 divisions |
Delay between any Two Channels (Typical) | ≤100 ps (50 Ω, DC coupling and equal V/div at or above 10 mV/div) |
Offset Range |
Range | 1MΩ | 50Ω |
1mV/div to 50mV/div | ±1V | ±1V |
50.5 mV/div to 99.5 mV/div | ±0.5V | ±0.5V |
100 mV/div to 500 mV/div | ±10V | ±10V |
505 mV/div to 995 mV/div | ±5V | ±5V |
1 V/div to 5 V/div | ±100V | ±5V |
5.05 V/div to 10 V/div | ±50V | n/a |
Offset Accuracy | ±(0.005 x |offset – position| + DC Balance) Note: Both position and constant offset term must be converted to volts by multiplying by the appropriate volts/div term |
Channel-to-Channel Isolation (Any two channels at equal Vertical Scale settings) (Typical) | ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated BW |
Horizontal System Analog Channels |
Maximum Sample Rate (All channels) | 5 GS/s |
Maximum Record Length (All channels) | Standard Config. | 12.5M points |
With Option 2RL | 25M |
With Option 5RL | 50M |
With Option 10RL | 125M |
Maximum Equivalent Time Sampling Rate | 400 GS/s |
Maximum Duration at Highest Sample Rate | 25 ms |
Time Base Range | 250 ps/div to 1000 s/div |
Time Resolution (in ET/IT mode) | 2.5 ps/div |
Time Base Delay Time Range | –5 divisions to 5000 s |
Channel-to-Channel Deskew Range | ±75 ns |
Time Base Accuracy | ±5 ppm over any ≥1 ms interval |
Trigger Jitter (RMS) | ≤10 psRMS for Edge trigger type ≤100 psRMS for all Non-edge trigger types |