世界上唯一专门为了测量特殊效果颜料金属颜料、珠光颜料等的颜色而设计的台式多角度分光测色仪;颜料颜色控制的系统;10种入射角度与观测角度的组合,分别是25/170、25/140、45/150、45/120、45/90、45/60、45/25、75/120、75/90;广泛适用于涂料、油漆、汽车、化妆品、油墨等行业。
FX10 Multiangle Spectrophotometer FeaturingIncluding:
· Ten sets of illuminating/viewing angles
· The ability to measure second-generation interference pigments.
· Conformance to DIN 6175-2 Farbtoleranzen für Automobillackierungen -- Teil 2: Effektlackierungen Tolerances for automotive paints – Part 2: Goniochromatic paints
· Conformance to ASTM E2194 Standard Practice for Multiangle Color Measurement of Metal Flake Pigmented Materials
· Ruggedeasy to use -- can be used in production laboratories
· Each illuminator/receiver pair are machined to ensure precise angular geometry
· Each illuminator/receiver pair is optically coupled to a single lampa single spectral analyzer
· Warranty - One year partslabor return to an authorized Datacolor service center
FX10 Specifications:
· Illuminatingviewing angles:
Illuminating Angle
Viewing Angle
Aspecular Angle
25°
170° -15°
25°
140° +15°
45°
150° -15°
45°
120° +15°
75°
120° -15°
75°
90° +15°
45°
110° +25°
45°
90° +45°
45°
60° +75°
45°
25° +110°
· Wavelength range - 400 to 780 nm
· Data output 10 nm at 10 nm effective bandpass
· Photometric range of 0 to 1000%
· Repeatability at all angles - average of 0.08maximum of 0.15 CIELAB Color Difference Units on a Spectralon plaque
· Inter-instrument agreement based at 45°/90° readings of 12 BCRA tiles environmental conditions apply - 0.2 CIELAB Color Difference Units average for BCRA tile set0.5 CIELAB Color Difference Units maximum for BCRA tile set
· Sample Aperture - 22 mm 0.9in wide x 69 mm 2.7 in long
· RS232 serial communication interface
· Size - 490 mm 19.3 in wide x 450 mm 17.7 in deep x 320 mm 12.6 in high
· Weight - 32.4 kg 71.5 lb
· Power - 110 to 250 VAC; 50 to 60 Hz; 240 VA maximum, 140 VA typical
FX10 QC Software Package FeaturingIncluding:
· Color spaces: CIELAB Lab, CIELAB LCh, CIELUV, XYZ, xyY
· Illuminants: A, C, D50, D55, D65, D75, F2, F7, F11, equal energy 100 at all wavelengths
· Observers: 2°, 10°
· Color difference equations: CIELAB, CMCl:c, DIN 6175-2 for both manufacturersusers of coatings
· User-based security system
· Data stored in an ASCII format readable by ExcelWord