Model | 19071 | 19073 |
功能 | ACV | ACV/DCV/IR |
Withstanding Voltage Test 耐压测试 | Output Voltage | AC : 0.05 ~ 5kV | AC : 0.05˜ 5kV, DC : 0.05 ~ 6kV |
Load Regulation | 1% + 5V |
Voltage Resolution | 2 V |
Voltage Accuracy | 1% + 5counts |
Cutoff Current | AC : 0.1mA~ 20mA | AC : 0.1mA ~ 20mA, DC : 0.01mA ~ 5mA |
Current Resolution | AC : 1μA, DC : 0.1μA |
Current Accuracy | 1.5% + 5counts |
Current Frequency | 50Hz/ 60Hz |
Test Time | 0.1~999 sec, continue |
Ramp up Time | 0.1~999 sec, off |
Waveform | Sine wave |
Insulation Resistance Test 绝缘测试 | Output Voltage | - | DC : 50 ~ 1000 V |
Voltage Resolution | - | 2V |
Voltage Accuracy | - | 5% + 5counts |
IR Range | - | 0.1MΩ ~ 50GΩ ≧ 500V : 1MΩ ~ 1000MΩ ± (4 % + 5 counts), 1GΩ ~ 10GΩ± (7 % + 5 counts), |
Resistance Accuracy | - | 10GΩ ~50GΩ± (12 % + 5 counts), |
< 500V : 0.1MΩ ~1000MΩ ± (7 % + 5 counts) |
< 100V : 0.1MΩ ~1000MΩ ± (10 % + 5 counts) |
ARC Detection | Setting Mode | Programmable setting |
Detection Current | AC : 1mA ~ 15mA, DC : 1mA ~ 5mA |
Minimum Pulse Width | 10 μs approx. |
Secure Protection Function | Fast Output Cut-off | Approx. 0.4mS, after NG happen |
Fast Discharge | Approx. 0.2S, Typical |
Ground Fault Interrupt | 0.5mA ± 0.25mAac (ON), OFF |
Continuity Check | 0.1Ω ~ 5.0Ω ± 0.2Ω, GC MODE |
Panel Operation Lock | Yes |
GO/NG Judgment Window | Indication, Alarm | GO: Short sound; NG: Long sound |
Data Hold | Least tests data memories |
Step Hold | Step signal trigger ON / OFF |
Memory Storage | 10 tests setups with 60 steps pre setup |
General | Operation Environment | Temperature: 0?C ~ 40 ?C, Humidity: ≦ 80 % RH |
Power Consumption | No load : <60 W, With rated load : ≦ 300 W |
Power Requirement | 100V / 120V / 220V / 240V, 50 / 60 Hz |
Dimension (H x W x D) | 105x272.8x350mm/ 4.13x10.74x13.78inch | 105x270x350mm/ 4.13x10.63x13.78inch |
Weight | 11 kg / 24.23 lbs |
Certification | UL, TUV, CE |