AC WITHSTAND VOLTAGE |
---|
Output Rating |
5KVAC / 30mA | 5KVAC / 30mA | 5KVAC / 40mA | 5KVAC / 40mA |
---|
| Range | Resolution | Accuracy |
Output Voltage,ACV |
0 - 5000 | 1 | ±(2% of setting + 5V) |
---|
Output Frequency |
50Hz / 60Hz ± 0.1%, User Selectable |
---|
Output Waveform |
Sine Wave ,THD. < 2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
---|
Output Regulation |
± (1% of output + 5V), From no load to full load |
---|
SETTINGS |
---|
Hi-Limit AC Current, mA |
0 - 30.00 | 0 - 30.00 | 0 - 40.00 | 0 - 40.00 |
---|
Resolution : 0.01, Accuracy : ±(2% of setting + 3counts) |
Lo-Limit AC Current, mA |
Range : 0 - 9.999 |
---|
Resolution : 0.001, Accuracy : ±(2% of setting + 3counts) |
Ramp Time, second |
0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) |
---|
Dwell Time, second |
0, 0.3 - 999.9
(0=continuous) | 0.1 |
---|
Arc Detection |
The range is from 1 - 9 (9 is the most sensitive) |
---|
AC Current Offset, mA |
0 - 2.000 or Auto Set |
---|
DC WITHSTAND VOLTAGE |
---|
Output Rating |
6KVDC/9999µA |
---|
Output Voltage, DCV |
0 - 6000 | 1 | ±(2% of setting + 5V) |
---|
Output Ripple |
< 5% (6KV / 9999 μA at Resistive Load) |
---|
SETTINGS |
---|
Hi-Limit DC Current, µA |
0 - 9999 | 1 | ±(2% of setting + 3counts) |
---|
Lo-Limit DC Current, µA |
0 - 999.9 | 0.1 | ±(2% of setting + 3counts) |
---|
Ramp Time, second |
0.1 - 999.9 | 0.1 | ± (0.1% of setting + 0.05s) |
---|
Dwell Time, second |
0, 0.4 - 999.9
(0=continuous) | 0.1 |
---|
Arc Detection |
The range is from 1 - 9 (9 is the most sensitive) |
---|
DC Current Offset, ?A |
0 - 200.0 or Auto Set |
---|
Ramp High Current |
ON/OFF, User Selectable |
---|
Charge - Low Current, µA |
0 - 350.0 or Auto Set |
---|
Discharge Time |
≦ 200 msec |
---|
INSULATION RESISTANCE |
---|
Output Rating |
1KVDC / 9999MΩ |
---|
Output Voltage, DCV |
100 - 1000 | 1 | ± (2% of setting + 3V) |
---|
Hi-Limit Resistance, MΩ |
0, 1 - 9999 (0=OFF) | 1 | 100 - 499V
1 - 999.9, ± (8% of setting + 2counts)
500 - 1000V
1 - 999.9, ± (2% of setting + 2counts)
1000 - 9999, ± (8% of setting + 2counts) |
---|
Lo-Limit Resistance, MΩ |
1 - 9999 | 1 |
---|
Delay Time, second |
0, 0.5 - 999.9
(0=continuous) | 0.1 | ±(0.1% + 0.05sec) |
---|
Charge - Low Current, µA |
0 - 3.500 or Auto Set |
---|
GROUND BOND (7440 ONLY) |
---|
Output Rating |
30A / 600mΩ, 8V |
---|
Output AC Current, A |
3.00 - 30.00 | 0.01 | ± (2% of setting +3 counts) |
---|
Output AC Voltage, V |
3.00 - 8.00 | 0.01 | ± (2% of setting +3 counts) at Open circuit |
---|
Output Frequency |
50Hz/60Hz ±100ppm, User Selectable |
---|
SETTINGS |
---|
Hi-Limit Resistance, mΩ |
0 - 150 (3 - 30A) | 1 | ± (2% of setting +2 counts) |
---|
Lo-Limit Resistance, mΩ |
0 - 600 (3 - 10A) |
---|
Lead Resistance Offset, mΩ |
0 -200 | 1 | ± (2% of setting +2 counts) |
---|
Dwell Time, second |
0, 0.5 - 999.9 sec (0=continuous) | 0.1 | ± (0.1% of setting + 0.05s) |
---|
MEASUREMENT |
---|
AC/DC Voltage, KV |
0 - 6.00 | 0.01 | ±(1.5% of reading ) > 500V
±(1.5% of reading + 1 count) < 500V |
---|
DC Voltage, V (IR) |
100 - 1000 | 1 | ±(1.5% of reading + 3V) |
---|
AC Current Range, mA |
0 - 3.500 | 0 - 3.500 | 0 - 3.500 | 0 - 3.500 |
---|
3.00 - 30.00 | 3.00 - 30.00 | 3.00 - 40.00 | 3.00 - 40.00 |
AC Current Resolution, mA |
0.001 / 0.01 |
---|
AC Current Accuracy, mA |
± (2% of reading + 3counts) |
---|
DC Current Range, µA |
0 - 350.0 | 0.1 | ± (2% of reading + 3counts) |
---|
300 - 3500 | 1 |
3000 - 9999 | 10 |
AC Current, A (GB) |
0 - 35.00 | 0.01 | ±(2% of reading+3counts) |
---|
Resistance, , MΩ(IR) |
1 - 9999
(Auto Range) | 0.001
0.01
0.1
1 | 100 - 499V
1 - 999.9, ± (8% of reading + 2counts)
500 - 1000V
1 - 999.9, ± (2% of reading + 2counts)
1000 - 9999, ± (8% of reading + 2counts) |
---|
Resistance, , mΩ (GB) |
0 - 600 | 1 | ±(2% of reading+2counts) |
---|
GENERAL |
---|
Input Voltage AC |
115V / 230Vac ± 15%, 50Hz / 60Hz ± 5%, max. current for 7410 model, max. current 6.3A for 7420 / 7430 / 7440 model |
---|
PLC Remote Control |
Input : Test, Reset, Memory1, 2, 3 |
---|
Output : Pass, Fail, Processing |
Memory |
50 memories, 8 steps/memory |
---|
Display |
20×2 LCD with back light |
---|
Key Lock |
To prevent unauthorized alteration of the test Parameters |
---|
Calibration |
Build-in software and external calibrated meters |
---|
Alarm Volume Setting |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
---|
Rear Panel Output |
YES | YES | YES | YES |
---|
Built-in Scanner |
- | 8 CHANNELS | OPTION | OPTION |
---|
External Scanner |
YES | YES | YES | YES |
---|
Smart GFI |
YES | YES | YES | YES |
---|
Analog Meter |
- | - | YES | YES |
---|
RS232 Interface |
OPTION | OPTION | YES | YES |
---|
GPIB Interface |
OPTION | OPTION | OPTION | OPTION |
---|
Printer Interface |
OPTION | OPTION | OPTION | OPTION |
---|
TUV-GS listed |
- | - | YES | YES |
---|
CE Compliance |
- | - | YES | YES |
---|
Environment |
0-40℃, 20-80%RH |
---|
Dimension (W × H × D), mm |
430 × 133 × 300 | 430 × 133 × 500 | 430 × 133 × 500 |
---|
Net Weight |
12Kg | 18Kg | 22Kg |
---|
STANDARD ACCESSORIES |
---|
Power Cord(10A) |
×1 |
---|
Fuses |
×2 (Including a spare contained in the fuse holder) |
---|
Hipot Test Lead ,1.5m(1101) |
×1 |
---|
Return Test Cable(1102) |
×1 |
---|
Ground Bond Test Lead 30A, 1.5m (1103) |
- | - | - | ×1 |
---|
Ground Bond Return Lead 30A, 1.5m (1104) |
- | - | - | ×1 |
---|
Hipot Output Link Lead, 1.5m (1109) |
- | ×8 | - | - |
---|
RS232 Link Cable 1.5m (1130) |
- | - | ×1 | ×1 |
---|
Disk (RS232 test program) |
- | - | ×1 | ×1 |
---|
*product specifications are subject to change without notice. |