Model
02PG
05PG
07PG
15PG
30PG
50PG
70PG
120PG
Unit
Recommended operating conditions
Pressure type | Gauge pressure | - |
Rated pressure | 13.79 | 34.47 | 48.26 | 103.4 | 206.8 | 344.7 | 482.6 | 827.4 | kPa |
Measurable pressure range | -13.79 ~13.79 | -34.47 ~34.47 | -48.26 ~48.26 | -98.07 ~103.4 | -98.07 ~206.8 | -98.07 ~344.7 | -98.07 ~482.6 | -98.07 ~827.4 | kPa |
Temperature range | 0~50 | °C |
Pressure media | Non-corrosive gases only(No liquid) | - |
Excitation current(Constant) | 1.5 | mADC |
Absolute maximum rating
Maximum load pressure | Twice of rated pressure | 1.5 times of rating pressure | - |
Maximum excitation current | 3.0 | mADC |
Operating temperature | -20~100 | °C |
Storage temperature | -40~120 | °C |
Operating humidity | 30~80(Non dew condition) | %RH |
Electric characteristics (Drive Current 1.5mA constant, ambient temperature Ta=25°C)
Output span voltage | 60~140(at 0kPa~rated pressure) | mV |
Offset voltage | ±20(at 0kPa) | mV |
Bridge resistance | 4000~6000 | Ω |
Response time | 2(for the reference) | msec |
Accuracy | TSO* | ±5 | %FS/ 0~50°C |
TCS* | 2.5 |
Linearity | ±0.5 | ±0.3 | ±0.5 | ±0.6 | %FS |
Pressure hysteresis | ±0.4 | ±0.2 | ±0.4 | %FS |