Model
|
19073
|
Test item
|
ACV/DCV/IR
|
Withstanding Voltage Test
|
Output Voltage
|
AC : 0.05˜ 5kV, DC : 0.05 ~ 6kV
|
Load Regulation
|
1% + 5V
|
Voltage Resolution
|
2 V
|
Voltage Accuracy
|
1% + 5counts
|
Cutoff Current
|
AC : 0.1mA ~ 20mA, DC : 0.01mA ~ 5mA
|
Current Resolution
|
AC : 1μA, DC : 0.1μA
|
Current Accuracy
|
1.5% + 5counts
|
Current Frequency
|
50Hz/ 60Hz
|
Test Time
|
0.1 ~ 999 sec, continue
|
Ramp up Time
|
0.1 ~ 999 sec, off
|
Waveform Sine
|
wave
|
Insulation Resistance Test
|
Output Voltage
|
DC : 50 ~ 1000 V
|
Voltage Resolution
|
2V
|
Voltage Accuracy
|
5% + 5counts
|
IR Range
|
0.1MΩ ~ 50GΩ
|
Resistance Accuracy
|
≧500V :1MΩ ~ 1000MΩ± (4 % + 5 counts),
1GΩ ~ 10GΩ± (7 % + 5 counts),
10GΩ ~ 50GΩ± (12 % + 5 counts),
< 500V : 0.1MΩ ~ 1000MΩ± (7 % + 5 counts)
< 100V : 0.1MΩ ~ 1000MΩ± (10 % + 5 counts)
|
ARC Detection
|
Setting Mode
|
Programmable setting
|
Detection Current
|
AC : 1mA ~ 15mA, DC : 1mA ~ 5mA
Minimum Pulse Width 10 μs approx.
|
Secure Protection Function
|
Fast Output Cut-off
|
Approx. 0.4mS, after NG happen
|
Fast Discharge
|
Approx. 0.2S, Typical
|
Ground Fault Interrupt
|
0.5mA ± 0.25mAac (ON), OFF
|
Continuity Check
|
0.1Ω ~ 5.0Ω ± 0.2Ω, GC MODE
|
Panel Operation
|
Lock Yes
|
GO/NG Judgment Window
|
Indication Alarm
|
GO: Short sound; NG: Long sound
|
Data Hold
|
Least tests data memories
|
Step Hold
|
Step signal trigger ON / OFF
|
Memory Storage
|
10 tests setups with 60 steps pre setup
|
General
|
Operation Environment
|
Temperature: 0 ~ 40℃
Humidity:≦80 % RH
|
Power Consumption No load
|
<60 W, With rated load :≦300 W
|
Power Requirement
|
100V / 120V / 220V / 240V, 50 / 60 Hz
|
Dimension (H x W x D)
|
105 x 272.8 x350 mm / 4.13 x 10.74 x 13.78 inch 105 x 270 x350 mm / 4.13 x 10.63 x 13.78 inch
|
Weight
|
11 kg / 24.23 lbs
|
Certification
|
UL, TUV, CE
|