CMASS公司专业临床质谱检测,帮您解答疑惑6PLUS1® Multilevel Serum Calibrator Set 25-OH-Vitamin D3/D2
Order no.: CMS-CH62039
- 25-OH-vitamin D2 25-羟基维生素D2
- 25-OH-vitamin D3 25-羟基维生素D3
MassChrom® 25-OH-Vitamin D3/D2 in Serum/Plasma
DESCRIPTION
6PLUS1® Multilevel Serum Calibrator Set, 25-OH-Vitamin D3/D2, lyophilised
These lyophilised calibrators from Chromsystems are based on human serum and should be used according to the instruction manual. They are designed to calibrate the measurement method in clinical diagnostics for the quantitative determination of 25-OH-Vitamin D3 and 25-OH-Vitamin D2 in serum/plasma. While following the instructions the calibrators should be handled and measured in the same manner as a patient specimen.
These calibrators are only available as multilevel calibrator set with 6 different concentration levels plus 1 blank value.
TECHNICAL DATAMethod of AnalysisLC-MS/MSContent7 x 1 ml (lyoph.)Parameter25-OH-vitamin D2, 25-OH-vitamin D3Concentrations
Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.
Substance | Unit | Calibrator 1 | Calibrator 2 | Calibrator 3 | Calibrator 4 | Calibrator 5 | Calibrator 6 | Blank Calibrator |
25-OH-Vitamin D3 | µg/l nmol/l | 9.1 22.6 | 19.0 47.5 | 33.8 84.2 | 66.0 165 | 105 261 | 141 353 | n.d. |
25-OH-Vitamin D2 | µg/l nmol/l | 4.7 11.4 | 14.0 33.9 | 28.5 69.1 | 56.1 136 | 102 247 | 144 348 | n.d. |
n.d. = not detected (< LOQ)
Reconstitution
Carefully remove the metal seal and rubber stopper from the vials. Add exactly 1.0 ml of distilled water into each vial. Return the stopper and incubate for 10 to 15 minutes at room temperature. Swirl the vials to dissolve the contents until homogeneity.
Storage conditions
Stored below -18 °C and unopened, the lyophilised calibrators are stable until the expiry date specified on the vial labels. The reconstituted calibrator can be stored in up to 1 weeks tightly capped at +2 to +8 C. If the calibrators cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 3 months). For more details about storage conditions please check your instruction manual.