Measurement Principle | UV Absorption at 254 nm; Dual Beam |
Federal Equivalent Method (FEM) | Yes, EQOA-0410-190 |
Linear Dynamic Range | 0 to 250 ppm |
Resolution | 0.1 ppb |
Precision (1σ rms noise) | Greater of 1.0 ppb or 2% of reading for 10-s average |
Accuracy | Greater of 1.0 ppb or 2% of reading |
Limit of Detection (2σ) | 2.0 ppb for 10-s average |
NIST-Traceable Calibration | Yes |
Measurement Interval | 2 s (Data averaging options: 10 s, 1 min, 5 min, 1 hr) |
Flow Rate (nominal) | ~1.8 Liter/min |
Flow Rate Requirement | >1.2 L/min |
Baseline Drift | < 1 ppb/day < 3 ppb/year |
Sensitivity Drift | < 1%/day < 3%/year |
Measurement Time, Frequency | 2 s, 0.5 Hz |
Response Time, of Step Change | 4 s, 2 points |
Measurement and Averaging Times | 2 s, 10 s, 1 min, 5 min, 1 hr |
Ozone Units | ppb, pphm, ppm, μg m-3, mg m-3 |
Pressure Units | mbar, torr |
Temperature Units | °C, K |
T and P Corrected | Yes |
Operating Temperature Range | 0 to 50 °C standard; optional -20 to 50 °C with low temperature modifications (pump and lamp heater cold-weather package) |
Operating Altitude Range | ~0-13.5 km (150-1,013 mbar) standard; optional ~0-25 km (30-1,013 mbar) with upgraded pressure sensor |
Power Requirement; Supplied by battery or 110/220 VAC Power Pack | 12 V dc or 120/240 V ac, nominally 650 mA at 12 V, ~8 watt (~7 watt with cell heater unplugged) |
Size | 3.7 x 8.5 x 11.5 inches (9.5 x 21.5 x 30 cm) |
Weight | 5.8 lb (2.6 kg); 3.1 lb (1.4 kg) without instrument case |
Internal Data Logger Capacity | 16,383 lines (2-s fast measurement mode = 9.1 hr; 10 s avg. = 1.9 days; 1 min avg = 11 days; 5 min avg = 1.9 mo; 1 hr avg = 1.9 yr |
SD Card Logger Capacity | > 2 GB (> 5-year capacity for 10-s averaging mode) |
Analog Inputs for Internal Logging of Other Instruments | 3 Analog Inputs, 0-2.5 V; For example, could log external T, P, and RH |
Data Outputs | RS232, 0-2.5 V Analog, LCD Display, External USB Converter (Optional 4-20 mA Current; request quote) |
Data Transfer Baud Rates | 2400, 4800, 19200 |
Output Ranges | User-Defined Scaling Factor in Menu |
DewLine™ | Yes |
Quiet, Long-Life Internal Air Pump | Yes |
Flow Meter | Yes |
Options | Battery; External Particle Filter; Serial-to-USB Adapter; GPS; Bluetooth; 4-20 mA Current Output; Cold-Weather Package (pump, lamp heater); High-Altitude Upgrade (pressure sensor); Extra-Long-Life, Low-Temperature Internal Air Pump; Low-Power Option; Rack-Mount Case Configuration |