INSULATION RESISTANCE |
| Range | Resolution | Accuracy |
Output Voltage, DCV | 30 – 1000 | 1 | ± (1% of setting + 1 V) |
Output Regulation DCV | ± (1% of output + 1 V) from no load to full load |
Output Ripple | < 2 % |
Discharge | Automatic Discharge of Device Under Test |
Indicator : Green < 30 V, Red > 30 V |
SETTINGS |
Low-Limit Resistance, MΩ | 0.1 – 999.9 | 0.1 | Same as Resistance Measurement |
1000 – 9999 | 1 |
10.0 GΩ– 50.0 GΩ for MODEL 8205 | 0.1 GΩ |
Delay Time, second | 1 - 999.9 | 0.1 | ±(0.1% + 0.05 sec) |
Dwell Time, second | 0, 0.1 – 999.9 (0=continuous) |
MEASUREMENT |
DC Voltage, V | 0.0–100.0 | 0.1 | ± (1% of reading + 1 V) |
101–1000 | 1 |
Resistance | 0.001MΩ - 50.0GΩ for MODEL 8205 0.001MΩ - 200.0GΩ for MODEL 8210, Auto Range | 0.001MΩ 0.01MΩ 0.1MΩ 1MΩ 0.01GΩ 0.1GΩ | 500–1000V |
0.1MΩ - 1GΩ,± (2% of reading + 2 counts) |
1GΩ - 20GΩ , ± (3% of reading + 2 counts) |
20GΩ –50GΩ,±(10% of reading + 2 counts) for MODEL 8205 |
20GΩ–200GΩ,±(10% of reading + 2counts) for MODEL 8210 |
30–499V |
0.1MΩ-1GΩ,± (3% of reading + 2 counts) |
1GΩ - 20GΩ , ± (5% of reading + 2 counts) |
GENERAL |
Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 1A/ Fast acting 250VAC |
PLC Remote Control | Inputs : Test , Reset, Interlock |
| Outputs : Pass, Fail , Progressing |
Display | 4 Digits LED Display |
Terminations | High Voltage Output – Alden Socket |
Shielded Return – BNC Connector |
Safety | Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS |
Calibration | Built-in software and external calibrated meters |
Environment | 0-40℃, 20-80%RH |
Dimension/Net Wight | 280mm(W)×89mm(H)×300mm(D)/ 4.5Kg |
STANDARD ACCESSORIES |
Power Cord(10A) | ×1 |
Fuses | ×2 (Including a spare contained in the fuse holder) |
Interlock Disable Key(1505) | ×1 |
Hipot Test Lead ,1.5m(1101) | ×1 |
BNC Connector Return Test Lead ,1.15m(1140) | ×1 |
*product specifications are subject to change without notice |