n
AH 1100/11A SPECIFICATIONS
$n
Range of capacitance values that can be ordered: Any value from 0.1 pF to 115 pF.
$n
Accuracy of initial setting: 2 + 1/C ppm at 1 kHz where “C” is the value of the 11A in pF.
$n
Stability in ppm per year: 0.3 + 1/C ppm/year
$n
Temperature coefficient relative to a change in ambient temperature: 0.01 ppm/°C
$n
Hysteresis from temperature cycling: 0.05 ppm
$n
Hysteresis from mechanical shock: 0.05 ppm
$n
AC voltage coefficient: 0.003 ppm/volt rms at 1 kHz
$n
DC voltage coefficient: 0.0001 ppm/volt
$n
Sensitivity to power line voltage changes: 0.0003 ppm per 1% change in power line voltage
$n
Dissipation factor: less than 0.000 003 tan delta
$n
Maximum allowable applied voltage: 250 volts peak
$n
Warm up time from power-on: 30 minutes
$n
Operating temperature range: 10° to 40° C
$n
Storage temperature range: -40° to +75° C
$n
Humidity: 0 to 85% relative humidity, non-condensing
$n
Power requirements: 40 watts max. during power-on, 20 watts after power-on
$n
Power frequency: 48 to 440 Hz
$n
Power voltage ranges: 85 to 115, 102 to 138, 187 to 253 and 204 to 276 volts rms
$n
AH 1100 packaging: The AH 1100 frame is 3.5 inches (8.9 cm) high and 15 inches (38.1 cm) deep behind the front panel. Hardware for rack mounting and a bail for bench top use are provided.
$n
Weight of AH 1100 only: 15 pounds (7 kg)
$n
AH 11A packaging: 4.45 wide by 8.0 long by 2.0 inches high (11.3 wide by 20.3 long by 5.1 cm high).
$n
AH 11A weight: 1.5 pounds (0.7 kg)
$n
Safety: Designed in accordance with UL1244, IEC348, and BS4743
$n
Radiated emissions: Designed to meet FCC and VDE class A requirements
$n
Warranty: The AH 1100 and AH 11A are covered by a three year warranty. Forward and return shipping is covered during the first three months of the warranty period.
$n
Note: Specifications are subject to change without notice.