简单介绍醋酸异戊酯检测管148 Isoamyl acetate,Measuring Range 10 to 200 ppm Number of Pump Strokes 2 Correction Factor 1 Sampling Time 4 minutes per pump stroke Detecting Limit 2 ppm (n=2) Color Change
简单介绍
醋酸异戊酯检测管148 Isoamyl acetate,Measuring Range 10 to 200 ppm Number of Pump Strokes 2 Correction Factor 1 Sampling Time 4 minutes per pump stroke Detecting Limit 2 ppm (n=2) Color Change
醋酸异戊酯检测管148 Isoamyl acetate的详细介绍
醋酸异戊酯检测管148 Isoamyl acetate
Isoamylacetate No.148CH3CO2(CH2)2CH(CH3)2(i-CH3CO2C5H11)
MeasuringRange | 10 to 200 ppm |
Number of PumpStrokes | 2 |
CorrectionFactor | 1 |
SamplingTime | 4 minutes per pumpstroke |
DetectingLimit | 2 ppm (n=2) |
ColorChange | Yellow → Paleblue |
ReactionPrinciple | CH3CO2(CH2)2CH(CH3)2 + Cr6+ + H2SO4 → Cr3+ |
Coefficient ofVariation | 15% (for 10 to 50ppm), 10% (for 50 to 200 ppm) |
ShelfLife | Temperature correction is necessary |
Correctionsfor temperature & humidity | 2 Years |
Store the tubes at cool and darkplace. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Alcohols | - | Plus error | Produce paleblue |
Ketones | - | Plus error | Produce paleblue |
Esters | - | Plus error | Produce paleblue |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | Diffusion tubemethod |
TLV-TWA | TLV-STEL | Explosive range |
100ppm | - | 1.0to 7.5% |